Oxygen Tank Sizes
The tank size determines how much oxygen a cylinder can hold. Each size has a specific "tank factor" used in duration calculations.
Medical Cylinders
- D Cylinder: Small portable cylinder (350 L capacity, factor: 0.16)
- E Cylinder: Standard portable cylinder (625 L capacity, factor: 0.28)
- M Cylinder: Medium stationary cylinder (3,000 L capacity, factor: 1.36)
- G Cylinder: Large stationary cylinder (5,300 L capacity, factor: 2.41)
- H/K Cylinder: Very large stationary cylinder (6,900 L capacity, factor: 3.14)
Aircraft Systems
- H125: Rotor wing aircraft system (2,166 L capacity, factor: 0.99)
- PC12: Fixed wing aircraft system (3,450 L capacity, factor: 1.56)
Tank factors are used to convert PSI readings into available oxygen volume.
PSI (Pounds per Square Inch)
PSI is the measurement of pressure inside the oxygen tank.
- Full Tank: Typically 2000-2200 PSI
- Standard Operating Range: 500-2000 PSI
- Reserve Level: Below 500 PSI (time to replace)
- Critical Level: Below 200 PSI (immediate replacement needed)
Always read the pressure gauge perpendicular to the gauge face for accurate readings.
Important: For safety reasons, never let your tank pressure drop below 200 PSI.
Always have a backup oxygen source available.
Flow Rate (Liters per Minute)
The flow rate determines how quickly oxygen is being delivered to the patient. Different clinical situations require different flow rates.
Common Flow Rates:
- 1-2 LPM: Low-flow oxygen therapy for COPD patients
- 2-4 LPM: Nasal cannula for mild hypoxemia
- 6-10 LPM: Moderate to severe hypoxemia
- 10-15 LPM: Non-rebreather mask for severe hypoxemia
- 15-25 LPM: Crisis situations, bag-valve-mask ventilation
- 25-60 LPM: High-flow oxygen therapy, critical patients
Higher flow rates will deplete your oxygen tank more quickly. Always match the flow rate to clinical needs.
Tank Duration Calculation Formula
The formula used to calculate how long an oxygen tank will last is:
Duration (minutes) = ((PSI - 200) × Tank Factor) ÷ Flow Rate
Example: For an E cylinder at 2000 PSI with a flow rate of 15 LPM:
Duration = ((2000 - 200) × 0.28) ÷ 15 = 1800 × 0.28 ÷ 15 = 504 ÷ 15 = 33.6 minutes
Understanding the Formula:
- PSI - 200 subtracts a safety margin of 200 PSI from the current pressure reading
- Tank Factor converts PSI to available oxygen volume (liters)
- Flow Rate is how many liters per minute are being delivered
The calculation gives you a realistic estimated duration that accounts for the 200 PSI safety margin.