Transvenous Cardiac Pacing

Updated Updated November 2026. What changed: The Precautions section has been removed and "Occlusion MI" is no longer listed as a contraindication; the indication now specifies symptomatic "(hemodynamically unstable)" arrhythmias. The Considerations on heparin after femoral vein placement and on Joint Commission alarm safety were also removed.
These protocols are here as an example, and are not to be used as an official resource or for patient care.

Indications:

  • Symptomatic (hemodynamically unstable) bradyarrythmias, tachyarrhythmias, and conduction system disturbances.
  • Transvenous pacing may be used as a bridge for those needing pacemaker revision or replacement.

Contraindications:

  • Aortic dissection, endocarditis, valve rupture, valvular pathology, prosthetic tricuspid valve.

Technique through Cordis:

  1. The CVC, either internal jugular or subclavian, with a cordis is in situ.
  2. The pacing wires have been sheathed, floated, and placed.
  3. The sending facility has initiated the transvenous pacing.
  4. TVP settings:
    • Most modern generators have a locking mechanism to reduce the risk of accidental changes. This is usually a button that looks like a key or a lock -- unlock this to change pacing settings.
    • Set the rate to maintain an appropriate cardiac output for the patient's needs.
    • There are two methods of setting current (measured in mA) to establish capture:
      • Start at 10-20 mA and decrease the current until capture is lost, then increase by 1 mA increments to maintain capture.
      • Set the output at 2 to 3 times the pacing threshold.
    • Sensing is measured in mV. If the patient has no intrinsic rate, the sensitivity is turned to the least sensitive value. If the patient has intrinsic activity, set the sensitivity to allow normal beats to be sensed and pacing to be inhibited. The sensitivity setting on the generator is decreased to allow for a greater sensitivity threshold (and vice versa).
  5. Ensure fluid is running through the cordis appropriately.
  6. Gather the appropriate equipment:
    • Generator
    • Spare batteries
    • Pacing wires
    • Return address of the sending hospital
  7. Ensure the locking mechanism (highlighted in yellow) is closed and mark the depth of the wires. Balloon port (highlighted in green).
  8. Ensure the pacemaker has electrical and mechanical capture.

Source page is a figure: photograph of a Medtronic temporary pulse generator connected by a cable to a transvenous pacing wire, with a balloon inflation syringe and a model heart at the tip of the wire; the locking mechanism on the wire is highlighted in yellow and the balloon port is circled in green.

Technique – Epicardial placement through chest wall

Source page is a figure: illustration of a chest with epicardial pacing wires exiting through the chest wall, labelled "Atrial wires" and "Ventricular wires", joined by "Pacing cables" to a "Dual chamber pulse generator".

Confirmation:

  • Chest X-Ray
  • 12 Lead EKG
  • Ultrasound

Troubleshooting:

  • Failure to Pace
    • Increase HR to 100 on the generator
    • If the pacing or sensing indicator flashes, assess the patient and check all connections and the battery. Check the position of the pacing electrode in the patient by chest X-ray.
    • If the pulse generator is turned on, but the indicators still are not flashing, change the battery. If that does not help, use a different pulse generator.
    • Check the settings if the pacemaker is firing too rapidly. If the pacing rate is correct or if increasing the sensitivity does not help, change the pulse generator.
  • Failure to sense
    • If the pacemaker is undersensing, increase the sensitivity by turning down the voltage (usually to 2 to 5 mV). If it is over sensing, decrease the sensitivity by turning up the voltage to 5 mV or higher.
    • Change the battery and then the pulse generator.
    • Remove items in the room that may be causing electromechanical interference, such as cell phones, headsets, and magnets. Check the ground wires on the patient's bed and other equipment for obvious damage. Unplug each piece and see whether the interference stops. When you locate the cause, notify the staff engineer and ask for it to be checked.
  • Failure to Capture
    • Increase output slightly; you can also increase sensitivity
    • Verify that the settings are correct.
    • If the heart is not responding, carefully check all connections and slowly increase the milli amperes as directed by your facility.
    • Check labs – potassium and magnesium, acidosis
    • Check the placement of transvenous wires
  • See Appendix for Medtronic pulse generator

Considerations:

  • Insulate exposed pacing leads when not in use and wear gloves whenever handling pacemaker electrodes to prevent microshock transmission of electrical current and the risk of serious arrhythmias.
  • If the manufacturer supplies a plastic cover with the pacemaker, place it over the pacemaker controls and setting lock feature to avoid an accidental setting change.
  • If atrial pacing is needed (typically for atrial arrhythmias with an intact atrioventricular [AV] node), a J-shaped pacing electrode catheter can be inserted through the superior vena cava into the right atrium until the electrode at the distal tip of the pacing catheter is lodged in the right atrial appendage.

Complications:

  • Micro-shock
  • Equipment failure (lead dislodgement and disconnection or catheter knotting)
  • Competitive or fatal arrhythmias
  • Pneumothorax
  • Hemothorax
  • Cardiac perforation
  • Cardiac tamponade
  • Diaphragmatic stimulation
  • Pulmonary embolism
  • Infection
  • Air embolism
  • Pain
  • Bleeding
  • Thrombophlebitis and lead displacement after threading the electrode through the femoral vein.

Documentation:

  • Reason for the temporary transvenous pacemaker
  • Current medication review findings
  • Laboratory test results
  • Pain scores
  • Medications administered
  • Confirmation of central venous access device placement via chest X-ray results
  • Date and time of pacemaker insertion
  • Time pacing started
  • Pacemaker settings
  • Type and location of the electrode
  • Vital signs
  • Cardiac rhythm strips from before, during, and after insertion if available
  • Any complications - interventions performed and response to those interventions
  • Confirmation of electrode placement - chest X-ray results and/or 12-lead ECG results
  • Post-procedure assessment