Monitoring Invasive Lines
Introduction: Critical care transport can involve monitoring many types of invasive lines, including arterial lines, Swan-Ganz catheters (PA and CVP), ICP (Intracranial Pressure) via bolt or external ventricular drain (EVD), and esophageal temperature probe.
Indications: Utilized in patients who are critically ill and allow for evaluation of immediate responses to therapy, assistance with medical diagnosis, and early identification of life-threatening conditions.
Technique:
Arterial lines:
- Continuous waveform monitoring and access to arterial blood for labs
- At risk of becoming occluded/kinked/dislodged during transport
- All arterial access lines should be transduced during transport
Swan-Ganz catheter:
- Can measure cardiac output episodically. Will continuously transduce CVP (blue port) and PA (yellow port) waveforms. The white lumen is for infusion; place a posi-flow cap on this port.
- When transporting patients with Swan-Ganz catheters, the distal tip (PA port) must be transduced and monitored to observe for spontaneous wedging. See below for ideal waveforms.
- Cardiac Output: normal value 4-8L/min, cardiac index 2-4L/min (indexed by BSA).
- PCWP (Pulmonary Capillary Wedge Pressure): normal range 8-12mmHg. Episodic reading, not continuous. Balloon MUST NOT be inflated for more than 5 seconds while measuring. Value reflects left ventricular end-diastolic pressure. Be aware that high PEEP (Positive End-Expiratory Pressure) values (>10cmH20) may raise wedge value. Wedge pressures should not be obtained without direct order of sending physician.
- PA Pressure: Must be monitored during transport to verify the balloon is not wedged.
- Systemic Vascular Resistance: Normal values 800-1200 dynes-sec/cm3. This is a calculated value and typically is not continuously measured.
- CVP: normal value 3-8mmHg. May be measured on IJ central line from distal lumen, or from blue port on Swan-Ganz
Source page is a figure: ideal Swan-Ganz waveforms as the catheter tip advances, with four heart diagrams (labelled A, RA, LA, PA, RV, LV) over a pressure tracing in mmHg (axis 0 to 30). Right Atrium: low tracing at about 0-5 mmHg. Right Ventricle: tall peaks from about 0 to about 25 mmHg. Pulmonary Artery: about 10 to about 25 mmHg with a dicrotic notch. Pulmonary Artery Wedge (drawn with the label "Inflated Balloon"): flat tracing at about 10 mmHg.
Temperature:
- Temperatures may be taken via esophageal temperature probe in intubated patients and by foley, temporally, orally, or rectally in non-intubated patients.
- The esophageal probe must be inserted in the distal esophagus to avoid the influence of the trachea on temperatures
Considerations:
- LifeLine monitors can monitor three pressure ports, and they should be used when clinically indicated and practical.
- This is extremely important in the care of the critically ill patient as it can guide response to therapy, help us recognize a deteriorating patient, and help recognize iatrogenic and non-iatrogenic complications
- Questionable values provided by our monitoring devices should always be followed up by clinical assessment and troubleshooting of the device.