Monitoring Invasive Lines

Updated Updated November 2026. What changed: The Swan-Ganz section adds a PA Pressure bullet (PA pressure must be monitored during transport to verify the balloon is not wedged) and a figure of ideal waveforms. Three statements were removed: ambient aircraft temperature monitoring on the Zoll per CAMTS, "monitoring is not a treatment", and the flight team understanding hemodynamic monitoring before leaving.
These protocols are here as an example, and are not to be used as an official resource or for patient care.

Introduction: Critical care transport can involve monitoring many types of invasive lines, including arterial lines, Swan-Ganz catheters (PA and CVP), ICP (Intracranial Pressure) via bolt or external ventricular drain (EVD), and esophageal temperature probe.

Indications: Utilized in patients who are critically ill and allow for evaluation of immediate responses to therapy, assistance with medical diagnosis, and early identification of life-threatening conditions.

Technique:

Arterial lines:

  • Continuous waveform monitoring and access to arterial blood for labs
  • At risk of becoming occluded/kinked/dislodged during transport
  • All arterial access lines should be transduced during transport

Swan-Ganz catheter:

  • Can measure cardiac output episodically. Will continuously transduce CVP (blue port) and PA (yellow port) waveforms. The white lumen is for infusion; place a posi-flow cap on this port.
  • When transporting patients with Swan-Ganz catheters, the distal tip (PA port) must be transduced and monitored to observe for spontaneous wedging. See below for ideal waveforms.
  • Cardiac Output: normal value 4-8L/min, cardiac index 2-4L/min (indexed by BSA).
  • PCWP (Pulmonary Capillary Wedge Pressure): normal range 8-12mmHg. Episodic reading, not continuous. Balloon MUST NOT be inflated for more than 5 seconds while measuring. Value reflects left ventricular end-diastolic pressure. Be aware that high PEEP (Positive End-Expiratory Pressure) values (>10cmH20) may raise wedge value. Wedge pressures should not be obtained without direct order of sending physician.
  • PA Pressure: Must be monitored during transport to verify the balloon is not wedged.
  • Systemic Vascular Resistance: Normal values 800-1200 dynes-sec/cm3. This is a calculated value and typically is not continuously measured.
  • CVP: normal value 3-8mmHg. May be measured on IJ central line from distal lumen, or from blue port on Swan-Ganz

Source page is a figure: ideal Swan-Ganz waveforms as the catheter tip advances, with four heart diagrams (labelled A, RA, LA, PA, RV, LV) over a pressure tracing in mmHg (axis 0 to 30). Right Atrium: low tracing at about 0-5 mmHg. Right Ventricle: tall peaks from about 0 to about 25 mmHg. Pulmonary Artery: about 10 to about 25 mmHg with a dicrotic notch. Pulmonary Artery Wedge (drawn with the label "Inflated Balloon"): flat tracing at about 10 mmHg.

Temperature:

  • Temperatures may be taken via esophageal temperature probe in intubated patients and by foley, temporally, orally, or rectally in non-intubated patients.
  • The esophageal probe must be inserted in the distal esophagus to avoid the influence of the trachea on temperatures

Considerations:

  • LifeLine monitors can monitor three pressure ports, and they should be used when clinically indicated and practical.
    • This is extremely important in the care of the critically ill patient as it can guide response to therapy, help us recognize a deteriorating patient, and help recognize iatrogenic and non-iatrogenic complications
  • Questionable values provided by our monitoring devices should always be followed up by clinical assessment and troubleshooting of the device.