Monitoring Cerebral Pressure
Introduction: Critical care transport can involve monitoring ICP (Intracranial Pressure) via bolt or external ventricular drain (EVD).
Indications: Utilized in patients who are critically ill patients and allow for evaluation of immediate responses to therapy, assistance with medical diagnosis, and early identification of life-threatening conditions.
ICP (Intracranial Pressure): Normal range <20mmHg.
EVD (ventriculostomy or extraventricular drain):
- Typically levelled at the tragus, which is a physiologic marker for the Foramen of Monroe
- Remember, an external ventricular drain can either be measuring or draining, but not both at the same time. If there is a waveform on the monitor, the EVD is measuring and clamped to drain (stopcock is closed). If there is a flat line on the monitor, the EVD is draining. If there is an ICP waveform on the monitor, then it is not draining. This is especially important when we are preparing to move the patient. Close the drain to move the patient and reopen when secured to the stretcher for transport.
- Clinically, we are unable to measure CBF (cerebral blood flow) but can use cerebral perfusion pressure (CPP) as a surrogate indicator of CBF.
- CPP = MAP - ICP, where ICP is the intracranial pressure obtained from an intracranial bolt or transduced EVD.
- It is recommended that the CPP be maintained at or greater than 60 mm Hg.
- The ICP may be reduced by facilitating venous drainage. This can be done by keeping the head in a neutral/midline position. Sedation, analgesia, and paralytics can also help reduce ICP. Lastly, it can also be reduced by drainage of CSF (Cerebrospinal Fluid) by ventriculostomy, followed by 3% saline, and lastly hyperventilation.
- Note color and amount of EVD drainage in patient's chart. Drain the buretrol prior to leaving the patient's bedside. This will give a more accurate account of volume out during transport. Assess ICP q 15 minutes. If there is a sudden change in color (bright red blood) or large amounts of drainage, notify receiving physician for orders and assess ICP every 5 minutes.
- Clamp EVD with movement and repositioning. After movement to the stretcher, re-level and open.
- Over draining can cause headaches, subdural hemorrhage, or herniation.
- If no drainage is noted, check patency by raising or lowering the drain, observing for pulsation or flow of CSF. Monitor ICP, as a clogged drain will result in increasing ICP. Notify the sending or receiving physician if the drain becomes clogged. If EVD is clogged (bloody CSF drainage), only a physician may flush the line or instill tPA into the catheter. Only a physician or midlevel provider may obtain a CSF specimen. Consider lowering the EVD set up to create a pressure gradient and clear the blockage, then re-level.
- Sending/Receiving Physician will order mmHg to set drain, and goal CSF drainage amount per hour.
- Assess site - no redness or leaking should be present.
- From report, gather the following and place in your documentation:
- Opening pressure
- Color of fluid
- ICP Target
- When to drain
- Max CSF to drain
If the electronic transducer on an External Ventricular Drain (EVD) is not working, you can still obtain a manual ICP reading using the following method:
- Leveling:
- Ensure the patient's head is properly positioned (usually at 30 degrees' elevation)
- Align the zero point on the EVD drainage system with the patient's external auditory meatus (ear canal)
- Prepare the system:
- Raise the level of the drainage system to its max by squeezing together the two tabs located directly above the drainage collection chamber and lifting (note: the EVD likely will not drain while at the max level)
- Leave the stopcock open to the patient
- Allow equilibration:
- Wait 30 seconds for the cerebrospinal fluid (CSF) to equilibrate in the catheter
- Observe and measure:
- Watch the fluid level in the clear catheter tubing
- The fluid will move to a level that corresponds to the patient's ICP
- Read the ICP value where the fluid level (meniscus) stabilizes on the mmHg or cm H2O scale printed on the EVD catheter
- Record the reading:
- Document the manual ICP measurement in mmHg or cm H2O
- Note the time of measurement and patient position
- Return the EVD to the prescribed drainage level by squeezing together the two tabs located directly above the drainage collection chamber and lowering it
Source page is a figure: two photographs of the EVD drainage system that sit in step 2 (Prepare the system). The first, after "Raise the level of the drainage system to its max", shows a hand at the drainage collection chamber beside the numbered pressure scale; the second, after "Leave the stopcock open to the patient", shows a hand holding the drainage system's scale, which is printed in mm Hg and cm H2O.
Bolt:
- Gives continuous ICP and no zeroing is required
- Do not bend the fiber optic catheter when moving and repositioning the patient. If this happens a physician will need to replace the fiber optic cable.
- If the patient has a Bolt and an EVD, ask the sending or receiving physician which reading they want to use to guide therapies.
Source page is a figure: ICP waveform tracings. The first shows a waveform with peaks labelled P1, P2 and P3 in descending height (described in the table below); the second, labelled "Poor Compliance", shows a rounded waveform with P1, P2 and P3 labelled at similar heights; the third, labelled "Dampened Waveform", shows a nearly flat tracing.
| ICP Waveform Component | Reflects |
|---|---|
| P1 - Percussion Wave | The ejection of blood from the heart transmitted through the choroid plexus in ventricles |
| P2 - Tidal Wave | Brain bulk or compliance, vasomotor paralysis, brain swelling or edema; reflects venous compartment. P2 = 80% P1 |
| P3 - Dicrotic Wave | Aortic valve closure |
Complications:
- Displacement - notify the receiving physician so they can be prepared to immediately place a new drain. Cover the site with a sterile, dry dressing.
- EVD's should never have a pressure bag attached to the transducer.
Considerations:
- LifeLine monitors can monitor 3 pressure ports and they should be used when clinically indicated and practical.
- This is extremely important in the care of the critically ill patient as it can guide response to therapy, help us recognize a deteriorating patient, and help transport teams recognize iatrogenic and non-iatrogenic complications
- Monitoring is not a treatment and should always be used in context with the patient's clinical appearance.
- Questionable values provided by our monitoring devices should always be followed up by clinical assessment and troubleshooting of the device.
- The flight team should also understand the basics of hemodynamic and intracranial pressure monitoring to stabilize the critically ill ICU patient before leaving.