High Flow Nasal Cannula

Updated Updated November 2026. What changed: HFNC is now considered failed at ≥ 60 L/min (adult) or ≥ 2 L/kg/min (pediatric) and over 60% FiO2, with a new line to consider converting to BiPAP before transport if the patient is close to those numbers. Cannula sizing is now a single rule (prongs fill 60% of the nares; was 80% for RAM cannulas), short transports are defined as typically under 60 minutes, and set-up photos were added.
These protocols are here as an example, and are not to be used as an official resource or for patient care.

Introduction: High-flow nasal cannula is a delivery method that administers high concentrations of oxygen through a ventilator circuit.

Indications:

  • Adults: hypoxic respiratory distress, pneumonia, cardiogenic pulmonary edema, DNR/DNI, post-extubation, pre-oxygenation prior to intubation
  • Pediatric: hypoxemic respiratory distress, bronchiolitis, pneumonia, croup, post-extubation, pre-oxygenation prior to intubation
  • Washes out dead space with oxygen, decreases preload, increases intrathoracic pressure, recruits alveoli, increases functional residual capacity, helps clear mucus
  • More comfortable and easier to tolerate than a mask for CPAP/BiPAP (Especially for pediatrics)

Contraindications:

  • Severe epistaxis
  • Suspected facial/skull fractures, cerebrospinal fluid leaks or tympanic membrane ruptures
  • Respiratory distress with PaCO2 >48mmHg on arterial blood gas, >51mmHg on venous blood gas
  • If an adult patient receiving ≥ 60L/min and over 60% FiO2, consider this therapy failed
  • If pediatric patient receiving ≥ 2L/kg/min and over 60% FiO2, consider this therapy failed
  • If the patient is close to the above metrics, consider converting to BiPAP prior to transport

Procedure:

  1. Place the patient on the monitor, recording HR, RR, BP, and oxygen saturation.
  2. Review the patient's history and evaluate contraindications or exclusions.
  3. Complete clinical assessment that should include the following:
    • Heart rate
    • Respiratory rate
    • SpO2
    • Work of breathing
    • Accessory muscle use
  4. Determine cannula size:
    • Prongs should fill 60% of the nares.
    • A clear gap should be visible around each prong when using on patients
    • Use the largest appropriate cannula for the patient
    • See specific equipment guides for further instruction

Necessary documentation:

  1. Cannula size
  2. Flow
  3. FiO2
  4. Temperature

Reassessment parameters:

  1. Changes in work of breathing, respiratory rate
  2. Disappearance of retractions, nasal flaring, grunting, tripod positioning
  3. Increase in SaO2, PaO2
  4. Increased mucus clearance
  5. Increase in overall comfort of patient
  6. Recheck arterial or venous blood gas

HFNC Set Up

  1. Use High Flow O2 Therapy Mode on the Hamilton Ventilator
  2. Note you will be given this notice:
  3. No apnea detection in this mode, and no disconnection detected. This is due to the one-way flow of oxygen. Perform a vent circuit test.
  4. Add a circuit extension set and connect to nasal cannula.
  5. Complete oxygen calculations to make sure we have enough oxygen to complete transport and 30 minutes beyond total transport time (bedside to bedside).
  6. Should you receive a high-pressure alarm at 50 mmHg, your nasal cannula size needs to be increased.
  7. The alarm below instructs the user to use nasal cannulas intended for high-flow oxygen delivery only.
  8. We will be transporting patients for short times (typically under 60 minutes) and humidification will not be used at this time.

Source page is a figure: photograph of the Hamilton ventilator screen showing the notice referred to in steps 2-3, reading "Hi Flow O2 therapy", "No apnea detection!", "No disconnection detection", with Flow: 17 l/min and Oxygen: 21 % displayed.

Source page is a figure: photograph of the Hamilton-T1 ventilator screen in HiFlowO2 mode showing the alarm referred to in step 7, reading "Use only interfaces intended for high flow O2. The use of unsuitable interfaces poses a risk to the patient. Active humidification is mandatory.", with Flow 15 l/min and Oxygen 50 % displayed and a "Loss of external power" banner at the top of the screen.

AIRVO Specific Information

Equipment:

  • Airvo module
  • Airvo heating canister and circuit
  • Oxygen flowmeter with high pressure tubing
  • O2 extension tubing
  • Appropriate size cannula

Initial Settings: choose correct mode for patient. To toggle between adult/pediatric modes: from main screen, hold down "play" button for 3 seconds. Butterflies appear on screen, indicating pediatric mode active.

  • Temperature: adjust as needed, initially set at 37C
  • FiO2: for ages 0-1 year, start at 40%. For ages 1-adult, start at 60%. Adjust as needed using the flowmeter, titrate to SpO2 goal
  • Flow: for pediatrics, start at 1.5L/kg/min; adjust up to 2L/kg/min. For adults, start at 60L/min, titrate down as tolerated
  • Cannula: size should occlude <50% of nare

Transport Mode: heater turns down to 31C, decreases humidification; all other settings remain the same. Saves battery power when on external battery pack. See Setup step 6 below for how to transition.

Source page is a figure: photograph of the Airvo screen in transport mode, showing a check mark, 25 L/min, 44 %O2 and a transport mode icon with a countdown reading 19:53.

Setup:

  1. Secure transport table to pram:
    • Secure UPS battery on table
    • Remove Airvo from pole, secure flowmeter to transport pram pole. Secure Airvo unit to table.
    • Plug Airvo unit into UPS battery GREY ports only
  2. At patient bedside, attach appropriate size cannula and turn unit on early, as the machine needs time to prime humidification chamber (will alarm if no cannula attached).
  3. Adjust initiation settings as needed for adult or pediatric patient (directions to switch mode above; this may be done early, and if turned off and back on, unit will remember last settings).
  4. Enable transport mode by pressing Up, Down, Silence buttons for 3-5 seconds, toggling over seven clicks to screen showing IV pole. Arrow up or down to choose image without red cross-out symbol, press Play button to save.
  5. When ready to transport patient, turn on transport mode by toggling to temperature screen, holding Silence button down for five seconds. Transport mode icon will appear and countdown of 20 minutes will appear.
  6. When in ambulance, immediately plug UPS battery into wall. Ensure generator is on. Turn off transport mode by toggling to Temperature screen, hold down Silence button for five seconds. Countdown should disappear and temperature value should reappear. Plug high pressure oxygen tubing into house oxygen.
  7. For transport of patient into receiving facility, repeat step 6.

Tips:

  1. Up and Down arrows held together will always Lock or Unlock screen
  2. Sideways arrow will toggle through screen options
  3. Up, Down, and Silence buttons held down together for 3-5 seconds will bring up advanced menu
  4. If unit NOT in transport mode and plugged into mobile battery, the battery will drain in less than 10 minutes