1066
Mechanical Ventilation
Updated
Updated November 2026.
What changed: Cuff pressure goal is now 20-30 mmHg (was 25-35), with pediatric cuffs targeting the lower end, and an EtCO2 strip must be documented at initial encounter, during transport and at the receiving facility. The Contraindications section and all Revel-specific content (90-degree swivel note, alarm-configuration settings) were removed, and charting now asks for maintenance of "sedation and analgesia" instead of "sedation/paralysis".
These protocols are here as an example, and are not to be used as an official resource or for patient care.
Indications:
Respiratory failure or cardiac arrest where the patient's own respiratory status is compromised and requires external means of ventilation, such as a bag-valve mask or mechanical ventilation.
Technique:
- Every patient will require different settings to maintain adequate oxygenation and ventilation. This is preferably guided by arterial blood gas values (may be venous blood gas). If unable to obtain blood gases, waveform capnography, oxygen saturations, and clinical assessment may guide settings.
- For adult patients, calculate Ideal Body Weight (IBW): Male = 50 + 2.3 (height in inches - 60); Female = 45.5 + 2.3 (height in inches - 60) and use to guide tidal volumes
Charting/QA/QI:
- Document ideal body weight
- Document 3 cufflator pressures taken, with goal between 20-30mmHg (pediatric cuff pressures should target the lower end of this range) and no air leak
- Document two methods of airway confirmation
- Document use of ventilator on all flights with intubated patients
- Document pPlat done, goal < 30mmHg. If not done, document reason for omission
- Document maintenance of sedation and analgesia
- Document EtCO2 strip for initial encounter, during transport, and at receiving facility
- Document if patient failed NPPV (includes CPAP and BiPAP and need for advanced airway during transport?
- Document if patient was transported with supraglottic airway.
- When charting ASV mode on Hamilton, initially chart PEEP (Positive End-Expiratory Pressure), FiO2, and % MV. After several minutes, chart Vent Monitoring with rate, VT, VTe, etc.
- Alarms: low pressure, high pressure, low minute ventilation (VE)
- Settings: Chart once at the beginning of call, re-chart if changed (power tool in ImageTrend)
| Vent mode | RR | I Time | O2 | Peep | PC | PS | TV |
|---|---|---|---|---|---|---|---|
| SIMV/PRVC | X | X | X | X | X | X | |
| AC/Volume | X | X | X | X | X | ||
| AC/Pressure | X | X | X | X | X | ||
| SIMV/Pressure | X | X | X | X | X | X | |
| BIPAP | X | X | X | X | X | ||
| CPAP | X | X | X |
Vent monitoring (ImageTrend): chart once and five minutes after any change to setting.
| Vent mode | RR | Vte | MV (Ve) | I: E | PIP | PPlat (at least once) |
|---|---|---|---|---|---|---|
| SIMV/PRVC | X | X | X | X | X | X |
| AC/Volume | X | X | X | X | X | X |
| AC/Pressure | X | X | X | X | X | X |
| BIPAP/CPAP | X | X | X | X | X | NA |
| SIMV/Pressure | X | X | X | X | X | X |
Considerations:
- Patients must be transported with an appropriately sized bag-valve mask in case of tube dislodgement.
- Be aware it is not possible to perform an inspiratory hold to measure plateau pressure while in a pressure-controlled setting
- Hamilton
- Pplat calculation. Driving pressure = VTE/ Cstat
- Pplat= driving pressure + peep
- Pplat calculation. Driving pressure = VTE/ Cstat
- Hamilton
- Isolation Patients- Filters should be placed on exhalation side of ventilator or closest to patient if single circuit, and exhalation side of ambu bag during manual ventilation
- In any volume setting, be sure to account for dead-space from equipment. This becomes especially important with pediatric patients, as their tidal volumes are smaller to begin with.
| Peak Inspiratory Pressure | 12 | 14 | 16 | 18 | 20 | 22 | 24 | 26 | 28 | 30 | 32 | 34 | 36 | 38 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 22mm (adult) | 17 | 20 | 22 | 25 | 28 | 31 | 34 | 36 | 39 | 42 | 45 | 48 | 50 | 53 |
| 15mm (peds) | 10 | 11 | 13 | 14 | 16 | 18 | 19 | 21 | 22 | 24 | 26 | 27 | 29 | 30 |
mL lost in circuit as dead space (table above).
Additional devices mL lost as dead space:
- ETCO2 devices: adult 6.6mL (for size 5.0 ETT and larger)
- Infant/neonate <0.5mL (for size 4.5 ETT and smaller)
- HME: adult 75mL, pediatric 30mL