Ventricular Assist Devices
Summary: A Ventricular Assist Device (VAD) is a mechanical device used to support circulation in a patient with significant cardiac ventricular dysfunction. The device is surgically implanted in the heart and provides continuous blood flow from left ventricle to aorta by a pump and graft.
LVAD patients can be identified by an electric driveline cable that comes directly out of their abdomen and connects to an external control pack powered by two external batteries they will be wearing with a bag, harness or vest.
The patient still has underlying heart function and rhythm that can be assessed and treated as appropriate per protocols. These patients are always on blood thinners, most likely Coumadin. (With INR usually 2-3)
Assess the patient:
Typically, LVAD patients have no discernible pulse. Blood pressure measurement requires manual BP cuff and Doppler or an arterial line. Typical goal "MAP" is 70-90mmHg, which on manual doppler assessment MAP will be the first heard doppler signal. See Monitoring Considerations below.
Other assessment parameters to define "stable":
- Level of consciousness
- Respiratory rate and work of breathing
- Signs of perfusion: skin color/temperature, capillary refill (HR >100 is hemodynamically unstable)
- Cardiac monitor, SpO2, blood glucose level, 12-lead EKG
Common VAD Complications:
- CVA
- TIA
- Arrhythmias
- Infections
- Sepsis
- Obstructions
- Pump Failure
- Bleeding
Charting Guidelines:
- Type of device (Heartmate II, III or Heartware)
- Flow, power, rpm every 15 mins
- Doppler MAP at same frequency as BP protocol
- Chart presence or absence of backup batteries
Treatment goal: maintain adequate end organ perfusion.
If transporting, notify destination of VAD patient inbound.
Is the patient alert and perfusing well?
- Address any non-LVAD related medical problems according to protocol
- Transport to University of Colorado Hospital for further treatment, if practical
- Contact VAD Coordinator
- Determine if VAD is running and functioning properly (auscultate chest for whirring sounds, examine VAD control unit for alarms)
- Assist ventilation as necessary
- Rule out other medical causes of altered mental status
VAD running: 250 mL bolus
VAD not running: Address VAD alarms/faults (power source?). If no power – connect to power source and restart.
Support as needed with Norepinephrine, Epinephrine infusions, blood products if meet criteria, RSI.
Reversing anticoagulation DIRECT ORDER FROM UNIVERSITY ONLY
No signs of perfusion?
- Initiate ACLS if VAD not running and unresponsive (PALS if patient prepubescent) and address underlying dysrhythmia or other problems per protocol
- Consider chest compressions if apneic with no clinical evidence of perfusion
- Consider defibrillation if arrhythmia and poor perfusion
Transport and Destination:
- It is vital to transport the patient's back-up batteries and emergency equipment with the patient.
- Unstable VAD patients should be transported to the nearest appropriate facility. University of Colorado Hospital is the only facility in the region that definitively treats VAD patients—and is therefore the preferred destination when patient condition is stable or is being seen at UCH and time permits with conditions/operational factors that allow transport.
- Contact VAD Coordinator as soon as possible at 24/7 pager # (303) 266-4522. For pediatric patients contact the Children's Hospital Colorado transplant coordinator pager at (303) 890-3503. Provide patient name, DOB, condition & ETA at destination for consultation and/or if transporting to University of Colorado Hospital. VAD coordinator will call back.
- VAD patient family members are excellent resources to assist with patient history and evaluation/repair of VAD alarms/faults.
- Device specific information for EMS can be found at: https://www.mylvad.com/medical-professionals/resource-library/ems-field-guides
Monitoring Considerations:
- Since there is non-pulsatile continuous flow, a pulse is usually not present and the BP must be obtained with a Doppler, usually from the brachial artery. Doppler BP is more reflective of a MAP and should be 70-90 mmHg.
- Place manual BP cuff on patient's arm
- Utilizing a doppler listen at the brachial artery for a "whooshing" sound inflate the cuff until it is no longer heard
- Deflate the cuff slowly, the MAP is the value when the "whooshing" sound returns
- Pulse oximetry monitoring may or may not be obtainable due to normal non-pulsatile state. Attempt to correlate O2Sat with ABG results if an arterial line is present however, UCH VAD coordinators would prefer to avoid unnecessary arterial punctures unless absolutely necessary due to these patients typically being anticoagulated. If unable to obtain O2sat then patients' oxygenation status should be assessed via color and level of consciousness and if in doubt treat patient with high flow oxygen and/or intubation if indicated per clinical assessment.
- Arterial lines are preferred for monitoring patient's overall perfusion state.
- Continuous waveform capnography (either inline if patient is intubated or via side stream nasal cannula) may be helpful in monitoring trends in patient's overall ventilation and perfusion status.
Management Pearls:
Patients are instructed to call 911 for:
- All major red alarms which are triggered when the flow < 2.5L/min which is usually caused by low preload, right heart failure, low batteries or drive line failure.
- If the AICD fires, which most patients have, this usually indicates VT/VF.
- If focal neurologic symptoms suggesting TIA or stroke occur
- If significant bleeding occurs and does not stop with application of pressure
- Due to the continuous flow, patients can remain fairly asymptomatic despite being in VT/VF/AF. They may only complain of vague symptoms like fatigue. If patient appears clinically stable, obtain 12 lead EKG and conservatively manage.
- The VAD is very preload dependent so "low flow" alarm may be triggered by dehydration
- Patients may be defibrillated, cardioverted and receive CPR if indicated.
- If the patient is unconscious follow the ACLS algorithm but look at the device controller to see if there is a red alarm. If there is a red alarm, then start CPR and follow ACLS resuscitation guidelines
Evaluation of LVAD patient with acute problem (uchealth VAD Patient Assessment):
- Assess responsiveness: if unresponsive, initiate emergency response & ACLS protocol; if responsive, obtain clinical history
- Physical examination: pulses, capillary refill, color & temperature of skin & extremities, auscultation of VAD motor sounds
- Device interrogation: look & listen for audible & visual alarms; connect patient to device monitor; document MCS parameters (pump flow / speed / power / pulsatility)
- Stable: HR less than 100 bpm, Doppler BP greater than 60 mmHg. Unstable: HR greater than 100 bpm, Doppler BP less than 60 mmHg
- Common LVAD complications: CVA/TIA (hold Coumadin, urgent head CT, stroke alert); Infection (obtain cultures, CXR, ABX per ID, MCS team will inspect sterile driveline site); Right heart failure (RV unsupported, IV fluids with caution, JVD?, edema?); other complications as applicable to patient's clinical H&P
- Differential diagnosis: Ventricular arrhythmias (IV fluids, reduce RPM, echo); Sepsis (IV fluids, pan cultures, CXR, CT scan, ABX per ID); GI bleed (IV fluids, type & cross for blood, hold Coumadin, GI consult); Flow obstruction (IV fluids, CXR, hemolytic panel, echo, heparin infusion); Pump failure (replace VAD controller, continue ACLS protocol)
Response to VAD patient with unresponsiveness or other altered mental status (uchealth VAD Patient Resuscitation):
- Assist ventilation if necessary and assess perfusion: normal skin color and temperature? Normal capillary refill?
- If perfusion is adequate: assess and treat non-LVAD causes for altered mental status, such as hypoxia, blood glucose, overdose, stroke
- If perfusion is not adequate: assess LVAD function (look/listen for alarms, listen for LVAD hum)
- If LVAD is functioning and MAP >50 mm Hg and/or PETCO2 >20 mm Hg: do not perform external chest compressions
- If LVAD is functioning and MAP/PETCO2 are below these thresholds: perform external chest compressions
- If LVAD is not functioning, attempt to restart LVAD: driveline connected? Power source connected? Need to replace system controller? If LVAD restarted, reassess MAP/PETCO2 as above; if LVAD cannot be restarted, perform external chest compressions
- Follow local EMS and ACLS protocols
- NOTE: VAD patients may be in a state of pseudo-PEA and not have a palpable pulse or recognizable BP yet have adequate perfusion. Signs of life and perfusion should be assessed by evaluation of: mental status, breathing, skin color and temperature, capillary refill
- Cardioversion and Defibrillation: VAD patients may present with ventricular arrhythmias and yet have adequate perfusion, IF the VAD is functioning. Consult Heart Failure or CT Surgery Attending MD for AWAKE VAD patients in V-Fib / V-Tach