Intra Aortic Balloon Pump (IABP)

These protocols are here as an example, and are not to be used as an official resource or for patient care.

Indications:

  • Refractory unstable angina
  • Severe multi-vessel or left main coronary artery disease
  • Cardiogenic shock with or without acute MI
  • Severe septic shock
  • Left ventricular failure

Contraindications:

  • Aortic Valve Insufficiency
  • Dissecting Aortic Aneurysm
  • Relative – Abdominal aortic aneurysm

Treatment goals: Reduce cardiac workload, increase cardiac output, improve myocardial oxygenation, increase coronary perfusion, maintain MAP >65mmHg on IABP console for organ perfusion.

See Cardiogenic Shock, other pertinent adult conditions.

IABP information:

  • Two different brands, Maquet and Arrow. Our pumps are Maquet so if transferring from Arrow to Maquet you will need appropriately sized adaptors for the helium driveline (size found at IABP insertion) and the fiberoptic will not be compatible
  • Battery life (Maquet and Arrow): 90 minutes. Battery life indicator has 5 lights, symbolizing ~20% battery life each
  • Maquet has two different size catheters, chosen based on patient height: 40cc (60-64in), 50cc (>64in)
  • Arrow has three sizes: 30cc (58-64in), 40cc (64-72in), 50cc (>72in)
  • IABP catheter usually placed through sheath in either femoral artery or axillary artery via cutdown+graft
  • Central lumen is 18 gauge, will clot easily. DO NOT USE FOR BLOOD DRAWS
  • The Datascope (Maquet) Clinical Support: 1-800-777-4222

Flight considerations:

  • Maquet rescue console, 2 batteries, and monitor weighs 53lbs, hospital cart weighs 61lbs, so together 114lbs
  • Pump will perform 1 autofill every 2 hours or with atmospheric pressure changes of 25-50mmHg, so every 1,000ft rise or 2,000ft drop in altitude
  • Turning pump on or a helium line disconnect will result in 6 autofills
  • Console contains enough helium for 36 auto fills; keep console in cart until loading for departure
  • Console should not operate in <50F or >104F, or above 12,000ft

Before leaving base on call: Bring IABP from base and red cardiac device bag. Can transport patient on current facility IABP console (MCR, Greeley, Longs Peak). Ensure adequate helium for duration of transport.

On scene at sending facility (and required charting):

  • Confirm placement via x-ray with sending provider, tip between 2nd-3rd intercostal space
  • Ensure patient on systemic anticoagulation (provider may defer if bleeding, must keep pump in 1:1 without anticoagulation)
  • Consider restraining/immobilizing affected extremity with knee immobilizer or soft restraint
  • Baseline assessment of affected extremity: color, temperature, pulses AND left radial pulse
    • Consider placing pulse oximeter on affected limb
  • Assess insertion site with sending RN
  • Ensure foley catheter placed to monitor urine output for IABP migration
  • Ensure central lumen connected to pressure line. If fiberoptic, transduce central lumen to Zoll. If not fiberoptic, must remain transduced to IABP console. Level and re-zero line once transitioned to Zoll or our IABP
  • If arterial sheath present, connect to pressure line and transduce to Zoll (not uncommon for this waveform to be over-dampened due to lines running through it). Level and re-zero line
  • Print strips from console (to change lines, press Preferences, select Printer, press Waveforms):
    • 1:2 ECG and arterial pressure, assess for timing problems
    • Arterial pressure and balloon pressure waveform

During transport:

  • Ensure console plugged into wall power
  • Do not position head of bed higher than 30 degrees
  • Reassess left radial pulse q30min, affected extremity as able
  • Use IABP console MAP to titrate medications
  • Print timing strips with change in patient condition i.e. heart rhythm

When dropping off at receiving facility:

  • Print strips from console:
    • 1:2 ECG and arterial pressure, assess for timing problems
    • Arterial pressure and balloon pressure waveform
  • Reassess insertion site with receiving RN

Switching IABP console:

  • Place consoles side by side. Power on console
  • Attach EKG cable and attach transport console EKG patches to patient, ensure tracing
  • If fiberoptic capable, remove fiberoptic key and attach to transport console (when fiberoptic removed, sending console will alarm and use EKG tracing to trigger). If not fiberoptic, switch arterial pressure cable to transport console at this time and zero pressure utilizing button on IABP console
  • Pause sending IABP console, switch driveline to transport console. When transport console recognizes driveline attached, press Start to begin pumping. May purge before consistently pumping
  • Remove sending facility monitoring equipment

Emergency procedures:

Cardiac arrest: unplug ECG cables (will force console to read off pressure, will sense CPR and assist)

Balloon rupture: noticed by bright red blood or dark red flecks in helium tubing

  1. Place pump in STANDBY to stop pumping
  2. Disconnect catheter from IABP console and pull air out of helium line
  3. Clamp helium tubing between Y fitting and male port

Console failure:

  • Detach catheter extender from IABP catheter
  • Attach 3-way stopcock & 60cc syringe to IABP catheters male luer fitting
  • Aspirate to ensure blood is not returned
  • Inflate IABP with 40cc air and immediately deflate – Repeat Q5min while inactive – notify receiving

Switching between Maquet-Arrow and vice versa:

  • Maquet to Arrow: Arrow has specific adaptors for helium driveline to connect into Arrow console, must match size of pump. Take end of Maquet helium line, insert nipple end of helium hose into blue (40cc) or orange (50cc) adapter plug, and plug into Arrow console
  • Arrow to Maquet: If on Arrow pump, need Maquet helium hose adaptor piece found in IABP bag. Utilize appropriately sized helium adapter hose and disconnect Arrow helium tubing at catheter insertion point at luer fitting. Attach Maquet helium adapter hose to luer fitting and insert nipple end into Maquet console. NOTE: do not throw out Arrow helium tubing, bring on transport in case receiving facility also uses Arrow to allow switch back

Alarms/troubleshooting:

  • Low Helium alarm: if less than 24 auto fills in cart, or less than 12 auto fills in transport console
  • Low Battery alarm: if less than 30 minutes of operating time, Q30second audible alarm, changes to high priority with <15min
  • Augmentation alarm: should be set 10mmHg below patients augmented pressure
  • Trigger alarm: related to poor R wave quality, adjust leads
  • Gas loss alarms: check helium circuit for leak/disconnection, check for blood in tubing. May manually purge circuit by pressing IAB Fill Key, then press Start (will stop pumping for 15 seconds, resume when purge complete). If blood in tubing: perform balloon rupture procedure
  • Fiberoptic alarms: IAB optical sensor failure: fiberoptic unable to read; clot on tip, not plugged in, defective. Check plug-in, attempt to recalibrate. Will postpone recalibration if patients MAP <65mmHg. If it does not resolve, ensure arterial pressure line from central lumen transduced to console

Modes:

  • Auto mode: default and most appropriate for all patients. Will select most appropriate lead & trigger. Deflation timing may still be manually adjusted. Best for paced rhythms
  • Semi-Auto mode: must manually choose trigger and adjust timing. If choosing "pacing", patient must be 100% paced. V/AV pacing will trigger on ventricular pacing spike, A pacing will trigger on R-wave. Never choose internal trigger.
Timing errorRecognitionResult
Early inflationIAB is inflated well before actual DN (aortic valve closure). Violates Rule 1 for inflation.Premature closure of aortic valve; reduces stroke volume/CO; increase in LVED volume; increase in LV wall tension
Late inflationDN is visible between points SYS/AUG. Violates Rule 1 for inflation.AUG less than optimum; decreased perfusion pressure and volume to coronary arteries
Early deflationASYS = SYS. Violates Rule 3 for deflation. May see "U" shape at ADIA.No afterload reduction
Late deflationADIA > DIA. Violates Rule 2 for deflation.Increased workload of left ventricle; increased MVO2

Normal balloon pressure waveform: zero baseline, IAB inflation, peak inflation (positive overshoot), plateau (full inflation of IAB), IAB deflation, peak deflation (negative overshoot), return to zero baseline.

Balloon pressure waveform tips:

  • Helium loss should be suspected if waveform doesn't return to baseline
  • If plateau becomes high or completely square at top, suspect high pressures/kinks in line/twisted catheter. If it is high at baseline after insertion, balloon catheter may be too large for aorta. Talk with sending provider about reducing amount of helium in balloon

CARDIOSAVE Rescue initial set-up:

  1. Press POWER button to turn IABP ON
  2. Connect ECG cable
  3. Connect fiber-optic sensor connector for Maquet/Datascope fiber-optic IAB, OR connect pressure cable for conventional IAB
  4. Connect IAB