Impella
Indications: severe cardiogenic shock unable to be stabilized by medications, less invasive therapies. Can be placed during interventions such as high-risk PCI or valvuloplasty. Can be bridge device to LVAD or transplant.
Impella device information:
- Impella is minimally invasive, micro-axial, non-pulsatile continuous flow pump that propels blood from left ventricle into aorta through a cannula, "percutaneous LVAD"
- Comprised of cannula inserted through Femoral Vein (Impella RP & RP Flex) or internal jugular vein (Impella RP Flex), console (Automated Impella Controller AIC), cassette, purge fluid circuit of dextrose containing fluid
- Two left sided devices CP (4.3 LPM max) & 5.5 (6 LPM max) & two right sided devices RP and RP Flex (>4 LPM)
- The 5.5 must be placed via arterial cut down
- If inserted properly and a thorough site assessment performed, the device should not malposition unless pulled which can cause patient to become very unstable
- One AIC weighs 26.8 lbs., if Bipella, will have 2 AIC
- Manufacturer is ABIOMED, representatives very clinically involved and available 24/7 for consult: 303-579-0694 or 800-422-8666
Contraindications to Impella:
- Mechanical aortic valve or heart constrictive device
- Presence of left ventricular thrombus
- Moderate to severe aortic valve stenosis
- Moderate or severe aortic valve insufficiency
- Echocardiographic assessment of aortic insufficiency graded as greater than or equal to 2+
- Severe peripheral arterial obstructive disease that would preclude Impella device placement
At bedside prior to transport:
- Confirm placement with sending physician with echocardiography (inlet 3.5cm from aortic valve for Impella CP, inlet 5cm from aortic valve for Impella 5.5 on parasternal long axis view), RP and RP Flex evaluated via CXR
- Confirm device plugged in (battery life is 60 minutes)
- Assess groin site and marking on catheter (in cm), tighten Tuohy lock to prevent catheter migration, confirm T Lock secure, confirm angle of insertion. For 5.5 only, ensure yellow pin is removed from the cath lock
- Obtain baseline settings:
- P level (P0-P9): Never <P2 for left, maintain flows >1.5 LPM on Right sided devices, unless on higher anti-coagulation and instructed by physician
- Flow (L/min, based on P level & patient preload/afterload)
- Aortic Placement Signal (mmHg, looks like aortic waveform)
- LV Placement Signal (mmHg, looks like a ventricular waveform)
- Purge Pressure (300-1100mmHg)
- Purge Infusion rate (2-30mL/hr, automatically adjusts to maintain purge pressure)
- Purge Solution (Dextrose in Water [D5W-D20W] with Heparin [25u/mL or Sodium Bicarb 25mEq/1L])
- Motor Current (Amps, pulsatile waveform indicates inlet and outflow are in different chambers=good placement. Flat indicates same chamber=poor placement)
- Maintain purge fluid bag upright to prevent air in purge tubing
- Obtain back up purge cassette from sending facility
- Immobilize affected limb using knee immobilizer or soft restraint on ankle
- If patient on systemic anticoagulation, maintain systemic as this is for the patient not pump
- Ensure foley catheter placed, make sure urine is clear (no discoloration/hematuria)
During transport:
- Use flowsheet to document all settings mentioned above q15min
- Maintain access to Impella plug/purge side arm
Switching consoles:
1. Prepare
- Confirm that the new AIC is powered on and ready
- Confirm patient is stable to be off support
2. Switch
- Remove the white connector cable from the original controller and plug it into the catheter plug on the front of the new controller
- Select OK within 10 seconds to resume at previous P level with gray selector knob when prompted (Impella will return to last known P level)
- Disconnect the yellow luer connector from the Impella Catheter to release the pressure in the purge cassette and immediately reconnect
- Transfer the purge cassette and purge solution from the original controller to the new controller
3. Confirm
- Patient is back on support with adequate flow
Proper Impella positioning — for left sided devices (Impella CP & 5.5):
- No active positioning alarms
- Pulsatile motor current
- Aortic Placement Signal – appearance of an arterial waveform
- LV placement signal (if at P4 or greater) – appearance of a ventricular waveform
For right sided devices (Impella RP and RP Flex):
- Pulsatile motor current
- Pulmonary Artery Placement signal (yellow) – appearance of a PA like waveform
Charting guidelines:
- Confirm placement on echo
- Laterality, size, cm marking at pick up and drop off, Tuohy secured, baseline settings, purge fluid & concentration
- Distal CMS of effected extremity
- Impella flowsheet
- Restraint/knee immobilizer to affected limb; HOB <30 if femoral
- Defib pads on pt.
- Color of urine (MUST HAVE FOLEY)
Potential complications:
Hematuria: caused by shearing of blood cells due to a malpositioned catheter
- Confirm placement, catheter external marking. Decrease P level by 1, assess preload and afterload
Suction alarms: caused by inadequate preload/blood to move from left ventricle into aorta. Can indicate hypovolemia or right heart failure
- Decrease P level by 1-2, consider 250-500ml saline bolus (goal CVP ~10 if able to measure), assess catheter still in good position. Once alarms stop, attempt to increase to prior P level
- If CVP >12 with persistent suction despite decreasing P level consider right heart support
Pump Position in Aorta or Ventricle: pump is malpositioned
- Turn pump to P-2, medically manage, and alert receiving
Air in Purge: red alarm and purge flow decreased to 0ml/hr
- Press PURGE MENU and select "De-Air Purge System"
- Select START to begin de-air process and follow on screen prompts
Low purge pressure alarms: viscosity of purge fluid not dense enough to clear circuit, or leak in system
- Check connections of purge tubing for leaks, check purge cassette for leaks. If none, consider increasing viscosity of dextrose in purge fluid (can go up to D20, rarely required)
High purge pressure alarms: viscosity of purge fluid too dense or kink in tubing
- Check tubing for kinks. If none, decrease viscosity of dextrose in purge fluid
- Ensure purge sidearm is clipped to white cable (CP, RP, RP Flex)
Cardiac arrest: follow routine Adult Pulseless Arrest protocol. Impella device will augment blood flow, likely will suction alarm due to lack of forward flow through heart.
- Decrease P level for CPR
- Do not touch Impella console during defibrillation/cardioversion
- Upon ROSC, if no positioning alarms return to previous P level, obtain ECHO if at facility. Assess for hematuria
Source page is a figure: Impella controller screen showing a red "Impella Position in Aorta" alarm at P-7: the Ao and LV placement signals both look like arterial waveforms (86/52 and 95/40 mmHg) and the motor current is flat (460/438 mA), with Impella flow 4.8 L/min, purge flow 2.5 ml/hr and purge pressure 301 mmHg. The on-screen alarm text reads: 1. Reduce the P-level to P-2 until imaging is available. 2. Reposition with imaging. 3. Return to preferred P-Level and confirm position with imaging.
Impella Position in Aorta:
- Will have red "Impella position in aorta" alarm
- Flat motor current
- Both Aortic placement signal and LV placement signal appear as aortic/arterial waveforms
- Follow management steps below
Source page is a figure: Impella controller screen showing a red "Impella Position in Ventricle" alarm at P-7: the Ao and LV placement signals both look like ventricular waveforms (101/12 and 101/12 mmHg) and the motor current is flat (460/438 mA), with Impella flow 4.8 L/min, purge flow 2.5 ml/hr and purge pressure 301 mmHg. The on-screen alarm text reads: 1. Reduce P-level to P-2. 2. When ready to reposition, press the MENU soft key then select Repositioning Guide.
Impella Position in Ventricle:
- Will have red "Impella position in ventricle" alarm
- Flat motor current
- Both Aortic placement signal and LV placement signal appear as ventricular waveforms
- Follow management steps below
Source page is a figure: Impella CP with SmartAssist controller screen with no alarm at P-9, labelled "Placement signal waveform" (aortic signal 110/75 mmHg over a ventricular LV signal 121/2 mmHg) and "Motor current waveform" (pulsatile, 997/852 mA), with Impella flow 3.8 L/min, purge flow 3.0 ml/hr, purge pressure 301 mmHg, cardiac output 5.0 L/min and cardiac power output 1.2 Watts.
Management of an improperly positioned pump:
- Reduce P level to P-2
- If still at sending facility request pump to be repositioned by ECHO
- If on transport medically manage patient and call receiving facility to update (caused by crew – fill out Baldwin)